[Models Supported] The RIGOL RSA3000/RSA5000 series spectrum analyzers, excluding RSA3000E models.
[Latest Revision Date] 19 Novermber 2025

[Notice] This firmware does NOT support RSA3015E, RSA3015E-TG, RSA3030E or RSA3030E-TG. For those models, visit the official website of RIGOL.

[Current Release]
00.03.08.00.02
    - Added remote‑command support for saving data to a USB flash drive.
    - Added SCPI commands to control Display → Web Control → On/Off settings.
    - Added timed self‑calibration at power‑on.
    - Added SCPI command support for retrieving raw IQ data in VSA mode.
    - Resolved an issue in GPSA mode where NDB bandwidth measurement only supported a single‑sided frequency‑response roll‑off.
    - Fixed incorrect save/load behavior for signal tables in EMI mode when amplitude units were changed.
    - Fixed an issue where VNA S21 measurements were overwritten after loading .trs files.
    - Fixed abnormal trace behavior caused by switching measurement modes in VNA S21 configuration.
    - Resolved SCPI trace‑saving failures in VSA mode.
    - Corrected out‑of‑tolerance measurement results for the 10 MHz resolution bandwidth.
    - Fixed incorrect Quasi‑Peak and Average limit lines in EMI mode for the EN 55022 standard.
    - Fixed preset‑mode labeling issues in VNA mode.

[Previous Releases]
00.03.07.04.01
    - Fixed an issue in EMI mode where multiple data points existed at the same limit line frequency and required a 1 Hz offset.

00.03.07.03.01
    - Corrected a 2 dB discrepancy between EMI mode (12 MHz adjacent measurement) and GPSA mode.

00.03.07.02.01
    - Added a 10× scan‑time extension in GPSA mode when the SPAN/RBW < 5, resolving spike artifacts in displayed traces.

00.03.07.01.01
    - Added storage and recall support for calibration data in VNA S11 and S21 measurements.

00.03.06.07.02
    - Added calibration‑data storage and recall for VNA S11 measurements.

00.03.06.06.02
    - Optimized RSA5000 scan‑time consistency relative to user‑configured settings.

00.03.06.05.02
    - Added persistence for limit lines and amplitude‑correction settings in power‑on configuration and state files.
    - Fixed incorrect return values for EMI limit line test command (:CALCulate:LLINe1:FAIL?).

00.03.06.05.01
    - Fixed Web Control touchscreen malfunction.
    - Updated system‑information section to display complete version numbers rather than 3‑segment format.


00.03.06 SP5 — 22 December 2023
    - Fixed incorrect return values of the EMI limit line test command (:CALCulate:LLINe1:FAIL?).
    - Resolved system freeze during remote‑command scanning.

00.03.06
    - Fixed system freeze when clearing and scanning in EMI mode.
    - Fixed inability to save .csv files in VSA mode.
    - Corrected readback errors through socket interfaces.
    - Resolved various SCPI command issues in EMI mode.
    - Added automatic detector switching to RMS Average during Adjacent Channel Power (ACP) measurements.
    - Added readback support for noise‑power values.
    - Enabled VNA measurement using SMA‑port calibration kits.
    - Fixed .csv file frequency resolution that previously only supported values to the thousand hertz level.

00.01.03
    - Resolved noise floor issues.
    - General software bug fixes.
    - Resolved multiple issues related to remote commands.

00.01.01
    - Fixed persistence display discontinuity issue.
    - Fixed RTSA noise floor issues.
    - Resolved remote‑command related issues.
    - Corrected problems related to options features.